UltraScan PRO Supplemental Manual for EasyMatch QC Ver. 2.0
32
Replacing the View Lamp
The view lamp is used to illuminate the sample when using the retroviewer feature. To replace the
lamp, obtain part number HL#A13-1004-022.
Figure 43. Retroviewer Lamp Part
To change the view lamp:
1.
Disconnect the UltraScan PRO from its power source.
2.
Wait two minutes for complete discharge of the capacitors.
3.
Disengage and remove the sample clamp.
4.
Loosen the two screws on the front cover with a flat-head screwdriver.
Figure 44. Front Cover Screws
Note: These screws are captive within the instrument cover housing and cannot be
removed
.
5.
Loosen the thumb screw on the inner panel bezel.