UltraScan PRO Supplemental Manual for EasyMatch QC Ver. 2.0
34
Figure 47. Replace View Assembly
8.
Plug the harnesses back into the front panel.
9.
Line the cover up with the guides on the bottom of the instrument.
10.
Replace the front cover on the instrument and tighten the two front screws as well as the
inner bezel thumbscrew.
Cleaning the Lens Surface and Didymium Filter
The lens and filter may be cleaned using photographic-quality lens solution and lens paper. Put a few
drops of solution on the lens paper and gently wipe the lens or filter in a circular motion for a few
seconds. Then wipe the lens or filter with a dry lens paper to remove streaks and any hazy film.
Cleaning the Sphere
Care should be taken to keep foreign materials from entering the sphere. The transmittance
compartment should be closed and the sample clamp placed against the reflectance port when the
instrument is not in use. The sphere is coated with Spectraflect
which can be damaged by putting
any objects into the sphere. If any foreign material (dust, lint, etc.) falls into the sphere, use the
following steps to remove the material.
1.
Move the sample clamp away from the reflectance port.
2.
Open the transmittance compartment and place a vacuum cleaner hose at the sphere
opening but do not insert the hose directly into the sphere. Cup your hand around the end
of the hose to provide a good seal.
3.
Quickly cover and uncover the reflectance port with your other hand. This creates gentle
air currents to swirl the foreign matter around and out of the sphere.
When shipping the UltraScan PRO, remove the sample clamp and tape a pad of foam rubber at the
reflectance port. Failure to do so may cause severe damage to the instrument.