SpectraTrend HT User’s Guide 3.2
61
measures height by detecting the angle of the light reflected to the sensor, height is not
measured absolutely, and all heights are relative. Therefore, the
MAX HEIGHT
distance
cannot be determined with a ruler. Instead, perform the
MAX HEIGHT
determination
procedure using the
DISTANCE SENSOR
in the
DIAGNOSTIC MENU
and enter the value
obtained as this parameter in the setup.
Figure 68. Distance and Height Parameters (Page 4)
PRODUCT SETUP, (TAB) PAGE 4: FILTER COEF
The
FILTER COEFF
function, when enabled, allows you to smooth the output data
without extending the update period. The output data may be smoothed by using the
TIME AVERAGE
read mode and a large number (such as 60 seconds). However, this
method could constrain the system to providing a reading only every 60 seconds, which
is unacceptable for some on-line processes. The
FILTER COEFF
function can provide the
same filtering effect as a long sample time while maintaining a worst-case sample time
of about one second.
The filter works by allowing you to select the ratio of old data to new data in each
update. The
FILTER COEFF
function performs the following transfer function on the
calculated color data prior to reporting:
VALUE_OUT = [NEW-VALUE X (1 - FILTER)] + OLD-VALUE X (FILTER)]
Enter an integer between 0 (off) and 90 (max effect).
Summary of Contents for SpectraTrend HT
Page 6: ...SpectraTrend HT User s Guide 3 2 6...
Page 21: ...SpectraTrend HT User s Guide 3 2 21...
Page 22: ...SpectraTrend HT User s Guide 3 2 22...
Page 23: ...SpectraTrend HT User s Guide 3 2 23...
Page 24: ...SpectraTrend HT User s Guide 3 2 24...
Page 25: ...SpectraTrend HT User s Guide 3 2 25...
Page 26: ...SpectraTrend HT User s Guide 3 2 26...
Page 27: ...SpectraTrend HT User s Guide 3 2 27...
Page 28: ...SpectraTrend HT User s Guide 3 2 28...
Page 76: ...SpectraTrend HT User s Guide 3 2 76...
Page 86: ...SpectraTrend HT User s Guide 3 2 86...