SpectraTrend HT User’s Guide 3.2
82
Figure 107. Self Test Diagnostic
The value of the checksum of the calibration data should remain constant unless the white tile’s
assigned reflectance is updated or the instrument is serviced. You should keep a record of the
checksum’s value. If it changes for unexplained reasons then the instrument may not function
correctly.
Figure 108. Firmware Versions & Check Sum Values
To return to the
DIAGNOSTICS MENU
, press the
LEFT
ARROW
key, to return to the
MAIN
MENU
, press the
RIGHT ARROW
key.