SpectraTrend HT User’s Guide 3.2
59
Choose
HITCH
when you wish to alter the readings made on the SpectraTrend HT to
better correlate to another color measurement instrument. A specific standard with
known color values from the other (reference) instrument will then be read with the
SpectraTrend HT and that reading manually adjusted within the product setup to match
the reference instrument. The modified standard is stored in the product setup for as
long as desired.
Figure 66. Select the Product Standard Type (Page 2)
PRODUCT SETUP, (TAB) PAGE 2: OFFSET CORRECTION
OFFSET CORRECTION
values may be entered for the color scale selected. The values
entered are added to the values read by the sensor.
PRODUCT SETUP, TAB: PAGE 2: TOLERANCES & WARNING LIMITS
Enter the desired
ALARM LIMIT
and
WARNING LIMIT
for each parameter into the
correct boxes.
CMC Parameters
Select
PAGE 3
to enter the
CMC/SHADE SORT
parameters.
Note that this applies only if the color index chosen is dE CMC. Shade Sorting Blocks
and Tolerance Band is not used with this equipment and may be ignored.
Summary of Contents for SpectraTrend HT
Page 6: ...SpectraTrend HT User s Guide 3 2 6...
Page 21: ...SpectraTrend HT User s Guide 3 2 21...
Page 22: ...SpectraTrend HT User s Guide 3 2 22...
Page 23: ...SpectraTrend HT User s Guide 3 2 23...
Page 24: ...SpectraTrend HT User s Guide 3 2 24...
Page 25: ...SpectraTrend HT User s Guide 3 2 25...
Page 26: ...SpectraTrend HT User s Guide 3 2 26...
Page 27: ...SpectraTrend HT User s Guide 3 2 27...
Page 28: ...SpectraTrend HT User s Guide 3 2 28...
Page 76: ...SpectraTrend HT User s Guide 3 2 76...
Page 86: ...SpectraTrend HT User s Guide 3 2 86...