SpectraTrend HT User’s Guide 3.2
63
Figure 0-70. External Triggering
PRODUCT SETUP, (TAB) PAGE 4: COLOR TRACKING
Choose whether
COLOR TRACKING
is enabled by setting the tolerances. When all
tolerance values are set to zero, the
COLOR TRACKING
feature is disabled. When
COLOR
TRACKING
is enabled, the product will be registered as present if the measured color is
within the preset tolerance of the target color. In this case, the number of valid color
readings per sample measurement interval will be displayed in the upper left hand
corner of the display for the current sample measurement. If the value of the valid
samples per measurement is zero, then no product is being measured.