successful, the versions of the firmware present would be displayed. Also shown is the value of
the checksum of the instrument’s calibration data as four hexadecimal digits.
Figure 85. Self-Test
The value of the checksum of the calibration data should remain constant unless the White
Tile
’s assigned reflectance is
updated, or the instrument is serviced. A
record of the checksum’s
value should be kept. If it changes for unexplained reasons, then the instrument may not
function correctly.
Figure 86. Self-Test Results
Press
DIAGS MENU
to return to the
DIAGNOSTICS MENU
. Press
MAIN MENU
to return to the
MAIN MENU
.