ColorFlex EZ Citrus Citrus
User’s Manual Version 2.0
12
Figure 7. Standardize
3.
Select
STANDARDIZE
from the ColorFlex EZ Citrus
’s main menu by moving the cursor
highlight to its position using the
UP
and
DOWN ARROW
buttons and then pressing
the
GO
button. The Standardization screen appears, prompting for the Black Glass.
Figure 8. Place Black Glass at Port
4.
Place the
Black Glass
on the measuring port with the shiny side of the glass against
the port. Check that the
Black Glass
is flat and in solid contact against the port and
that the white dot is facing forward towards the operator.
5.
Press the
GO
button. The ColorFlex EZ Citrus reads the glass and sets the instrument
bottom-of-scale. When it is finished, the screen prompts for the calibrated White
Tile.
Figure 9. White Tile Reading
6.
Replace the
Black Glass
with the calibrated
White Tile
from the standards box.
Figure 10. Confirmation of Standarization