ColorFlex EZ Citrus Citrus
User’s Manual Version 2.0
36
Figure 61. Setups Locked
AutoSave
Choose
YES
to save each measurement automatically to the instrument’s Datalog. Choose
NO
to save the measurements manually, only on demand. Each press of the
RIGHT ARROW
button
toggles between
YES
and
NO
.
Figure 62. AutoSave
Stdz Interval
Indicate the amount of time that must elapse (in hours) after the last standardization and
before the instrument prompts for re-standardization. Press the
RIGHT ARROW
button to
increase the number of hours or the
LEFT ARROW
button to decrease the number. Values
between 1 and 16 may be selected in increments of 1. Selecting 4 is suggested. The choice of
OFF
disables prompting for standardization, but its use is not recommended.
Figure 63. Standardization Interval
AutoSearch
One of the features of the ColorFlex EZ Citrus is that it can find the closest physical, numeric, or
hitch standard stored in the instrument’s memory to a sample read.
When
AUTOSEARCH
is highlighted and the
RIGHT ARROW
button is pressed, a submenu
appears. In this submenu, the following parameters may be set: