Service/maintenance
OPERATION MANUAL
Chapter 7
Minichiller®
V2.4.0en/09.08.21//1.30
52
7.3.2.1
Rinsing a thermofluid circuit with >Sight glass< [23]
PROCEDURE
Drain the temperature control unit.
→
»Draining with >Sight glass< [23]«
Residual thermal fluid can remain in the pump chamber and the internal lines after draining. Leave
the temperature control unit with open valves for a while.
Check the fill level in the collecting container. Observe the proper disposal of thermal fluid.
→
»Proper disposal of resources and consumables«
Re-fit the knurled screw to the
>Residues drain<
[10] (if present).
Re-fit the knurled screw to the
>Drain<
[8].
Connect the
>Circulation flow<
[1] with the
>Circulation return<
[2] at the temperature control
unit using a bypass hose.
Perform the following steps without attaching a short circuit hose, if the application used by you
(externally closed) is also dirty. In this case, leave your externally closed application connected to
the temperature control unit. This rinses the temperature control unit and your application at the
same time.
Fill
the system (minimum fill level) with the thermal fluid you wish to use.
→
Page 40, section
»Filling and venting with >Sight glass< [23]«
Vent
the system.
→
»Filling and venting with >Sight glass< [23]«
Adjust the
setpoint
and the cut-out value of the
overtemperature protection
to the thermal
fluid used.
→
and
→
Start
the
circulation
. The length of rinsing depends on the level of soiling.
Stop
the
circulation
.
Drain
the temperature control unit.
→
»Draining with >Sight glass< [23]«
Repeat the steps “Filling”, “Venting”, “Start/Stop circulation” and “Draining” until the drained
thermal fluid remains clear.
Remove the bypass hose after completely draining the temperature control unit.
Leave an application connected, if you have simultaneously rinsed a used application (externally
closed).
Leave the
>Drain<
[8] and
>Residues drain<
[10] (if any) open for a while to allow the thermal
fluid to evaporate in the temperature control unit.
Close the
>Drain<
[8] and
>Residues drain<
[10] (if any) once the thermal fluid has evaporated.
Remove the collecting container. Follow the instructions for the proper disposal.
→
Page 15,
section
»Proper disposal of resources and consumables«
Re-connect your application. (Only if you have rinsed the thermal fluid circuit using a bypass hose.)
Fill the temperature control unit with thermal fluid.
→
Vent the temperature control unit.
→
»Filling and venting with >Sight glass< [23]«
Restart the temperature control unit in normal mode.
7.3.2.2
Rinsing the thermofluid circuit with >Level indicator and drain< [38]
PROCEDURE
Drain the temperature control unit.
→
»Draining with >Level indicator and
Residual thermal fluid can remain in the pump chamber and in the internal lines after draining.
Therefore leave the temperature control unit open for a while.
Check the fill level in the collecting container. Follow the instructions for the disposal of thermal
fluid.
→
»Proper disposal of resources and consumables«
Reinsert the hose into the
>Level indicator and drain<
[38].
Connect the
>Circulation flow<
[1] with the
>Circulation return<
[2] at the temperature control
unit using a bypass hose.
Summary of Contents for MPC Minichiller 280
Page 1: ...MPC Minichiller...
Page 2: ......
Page 3: ...OPERATION MANUAL MPC Minichiller...
Page 4: ......
Page 10: ...OPERATION MANUAL Minichiller V2 4 0en 09 08 21 1 30 10...
Page 59: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 09 08 21 1 30 Minichiller 59 9 Annex...