Commissioning
OPERATION MANUAL
Chapter 2
Minichiller®
V2.4.0en/09.08.21//1.30
26
Side
Distance in cm (for operation in a tub)
Air cooling
Water cooling
[A1]
Top
Air outlet on top of unit: free standing
–
[A2]
Top
can be located under a bench
can be located under a bench
[B]
Left
min. 20
min. 20
[C]
Right
min. 20
min. 20
[D]
Front
min. 20
min. 20
[E]
Rear
min. 20
min. 20
2.3.1
EMC-specific notes
Connecting cables in general
Prerequisites for a failure-free operation of the temperature control units incl. their connections with external
applications: Installation and wiring must be carried out professionally. Related topics: "Electrical safety" and
"EMC-compliant wiring".
Cable lengths
For flexible/fixed cable routing of more than 3 meters, the following must amongst other things be observed:
- Equipotential bonding, grounding (see also technical data sheet "Electromagnetic compatibility EMC")
- Compliance with "external" and/or "internal" lightning/overvoltage protection.
- Design protection measures, professional cable selection (UV resistance, steel pipe protection, etc.)
Attention:
The operating company is responsible for compliance with national/international directives and laws. This also
includes the testing of the installation/wiring required by law or standards.
This device is suitable for operation in
“industrial electromagnetic environments”
. It meets the
“immunity requirements”
of the currently applicable
EN61326-1
, which are required for this envi-
ronment.
It also meets the
“interference emission requirements”
for this environment. It is a
Group 1
and
Class A
device according to the currently applicable
EN55011
.
Group 1
specifies that high frequency (HF) is only used for the function of the device.
Class A
defines
the interference emission limits to be observed.
2.4
Installation conditions
Temperature control unit is connected to the power supply line
DEATH FROM ELECTRICAL SHOCK BY DAMAGE TO THE POWER CABLE.
Do not put temperature control unit on power cable.
Summary of Contents for MPC Minichiller 280
Page 1: ...MPC Minichiller...
Page 2: ......
Page 3: ...OPERATION MANUAL MPC Minichiller...
Page 4: ......
Page 10: ...OPERATION MANUAL Minichiller V2 4 0en 09 08 21 1 30 10...
Page 59: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 09 08 21 1 30 Minichiller 59 9 Annex...