Service/maintenance
OPERATION MANUAL
Chapter 7
Minichiller®
V2.4.0en/09.08.21//1.30
50
PROCEDURE
Table-top models:
Turn off the temperature control unit.
Disconnect the temperature control unit from the power supply.
Close the customer’s shut-off valves in the cooling water supply and return lines.
Place a collecting container below the
>Cooling water inlet<
[13].
Remove the cooling water supply line and take out the hat-type strainer for inspection and
cleaning.
Clean the suction strainer under running water.
Following inspection/cleaning, reinsert the hat-type strainer and fasten the cooling water supply
line.
Remove the collecting container from below the
>Cooling water inlet<
[13].
Open the customer’s shut-off valves in the cooling water supply and return lines.
Connect the temperature control unit to the power supply.
Turn on the temperature control unit.
PROCEDURE
Free-standing models:
Switch off the temperature control unit.
Disconnect the temperature control unit from the power supply.
Close the customer’s isolating valves in the cooling water supply and return lines.
Remove the paneling around the cooling water supply [13], [14] and [15] (if present).
Place a collecting container below the
>Cooling water inlet<
[13] and another collecting contain-
er below the
>Cooling water drain<
[15] (if present).
Open the ball valve of the
>Cooling water drain<
[15] (if present). If the temperature control unit
is not equipped with a
>Cooling water drain<
[15]: Open the
>Cooling water outlet<
[13]. The
cooling water will begin to drain out. Allow the cooling water to fully drain.
Disconnect the
>Cooling water inlet<
[13] from the building’s cooling water supply. Located
immediately behind the
>Cooling water inlet<
[13] is the dirt trap.
Carefully detach the cover (hexagonal).
Remove the metal strainer located below.
Clean the metal strainer under running water.
Re-insert the metal strainer after cleaning work.
Carefully fasten the cover (hexagonal).
Connect the
>Cooling water inlet<
[13] to the building’s cooling water supply.
Close the ball valve of the
>Cooling water drain<
[15] (if present).
Remove the collecting containers below the
>Cooling water inlet<
[13] and below the
>Cooling
water drain<
[15] (if present). Dispose of the contents of the collecting containers. Follow the in-
structions for the proper disposal.
→
»Proper disposal of resources and con-
Mount the paneling around the cooling water supply [13], [14] and [15] (if present).
Open the customer’s isolating valves in the cooling water supply and return lines.
Connect the temperature control unit to the power supply.
Switch on the temperature control unit.
We are also happy to offer you service training. Please contact Customer Support.
→
Page 58,
7.3
Thermal fluid inspection, replacement and circuit cleaning
Observe the wiring diagram.
→
Extremely hot / cold surfaces, connections and thermal fluids
BURNS/FREEZING OF LIMBS
Surfaces, connections and tempered thermal fluids can be extremely hot or cold depending on
the operating mode.
Avoid direct contact with surfaces, connections and thermal fluids!
Wear your personnel protective equipment (e.g. temperature-resistant safety gloves, safety
goggles).