Setup mode
Chapter 4
OPERATION MANUAL
V2.4.0en/09.08.21//1.30
Minichiller®
39
4.1.5
Setting the setpoint
PROCEDURE
Switch on the temperature control unit.
Press and hold the
>SET key<
[D]. The setpoint is displayed.
Use the
>Arrow keys<
[C] to select the required value.
(up) the temperature value is increased, (down) the temperature value is decreased.
Release the
>SET key<
[D]. The new setpoint is set.
4.2
Filling, venting and draining
Observe the wiring diagram.
→
Extremely hot / cold surfaces, connections and thermal fluids
BURNS/FREEZING OF LIMBS
Surfaces, connections and tempered thermal fluids can be extremely hot or cold depending on
the operating mode.
Avoid direct contact with surfaces, connections and thermal fluids!
Wear your personnel protective equipment (e.g. temperature-resistant safety gloves, safety
goggles).
During an active circulation, the thermal fluid circuit is shut off by shut-off valves
MATERIAL DAMAGE TO THE CIRCULATING PUMPS INSTALLED IN THE TEMPERATURE CONTROL
UNIT
Do not close the thermal fluid circuit during an active circulation by means of shut-off valves.
Warm the thermal fluid to room temperature before stopping the circulation.
4.2.1
Filling and venting externally closed application
Non-compliance with the safety data sheet for the thermal fluid to be used
INJURIES
Risk of injury to the eyes, skin, respiratory tract.
The safety data sheet for the thermal fluid to be used must be read prior to using it and its
content must be respected.
Observe the local regulations/work instructions.
Wear your personal protective equipment (e.g. temperature-resistant safety gloves, safety
goggles, safety footwear).
Danger of slipping because floor and work area are contaminated. Clean the workplace; ob-
serve the proper disposal of thermal fluid and aids.
→
Semi-automatic venting
DAMAGE TO THE TEMPERATURE CONTROL UNIT
An elevated tolerance time for the pressure drop can damage the pump if the thermal fluid
level in the system is also too low.
Constantly observe the thermal fluid level on the
>Sight glass<
[23] or the
>Level indicator and
drain<
[38]. Fill up the thermal fluid during the venting phase so the thermal fluid level does
not fall below the minimum mark in the
>Sight glass<
[23] or the
>Level indicator and drain<
[38].
Calculate whether the capacity of the
>Expansion vessel<
[18] can absorb the expansion volume
during operation. Assume the following volumes for this calculation: [Minimum filling capacity of
the temperature control unit] + [Volume of the thermal fluid hoses] + [Jacket volume of your appli-
cation] + [10% / 100 K].
▪
During the fill process, ensure any necessary measures, such as earthing the tanks, funnels and
other aids, have been taken.
▪
Fill to the lowest possible height.