![HP HP 8924C Application Manual Download Page 94](http://html.mh-extra.com/html/hp/hp-8924c/hp-8924c_application-manual_166201094.webp)
94
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amrcvrt.fb
Chapter 3, CDMA Receiver Tests
Measuring Receiver Sensitivity and Dynamic Range
7. Arm a single measurement.
HP-IB Syntax
"TRIG:MODE:RETR SING" !selects Single measurement mode.
"TRIG:AST ‘ARM’" !starts the FER measurement.
"TRIG" !starts the FER measurement and other "active" measurements.
Manual Operation:
1. Position the cursor at the Arm field.
2. Press the knob to arm a measurement.
When a measurement is running, the Testing annunciator will be lit.
To Screen
CDMA CELLULAR MOBILE RECEIVER TEST
FER
Test Status
Display
Interim
Results
Yes/No
Max Frames
On/Off
Sector A
Power
-105.0
dBm/BW
%
Connected
Svc Opt 2
Testing
Passed
Failed
Max Frames
Errors Counted
Traffic
Data Mode
Svc Opt 2
Data Rate
Full
Eb/Nt
CDMA
CALL CNTL
Analog
RX TEST
Cnfig
PRNT CNFG
Frames Counted
Meas Cntl
Single/Cont
Arm
Disarm
Confidence
95.00
FER Spec
0.50
----
Traffic
-15.6
dB
AWGN
Off
%
%