![HP HP 8924C Application Manual Download Page 195](http://html.mh-extra.com/html/hp/hp-8924c/hp-8924c_application-manual_166201195.webp)
195
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amtxtst.fb
Chapter 4, CDMA Transmitter Tests
Measuring Access Probe Output Power
Cha
p
te
r 4
CDM
A T
ra
n
s
m
itte
r
Te
s
ts
Measurement Overview
1.
"Connect the test instruments as shown." on page 187
Connect the equipment.
MSUT and combiner.
2.
"Set the test parameter Îor ." on page 188
.
Set the test parameter Î
or
.
Screen: CDMA CALL CONTROL
Enter value in:
Sector A Power
3.
"Set the Test Set to ignore all access attempts." on page 189
.
Screen: CDMA CALL CONTROL
Enter value in:
Call Limit
4.
"Register the MSUT." on page 190
.
Screen: CDMA CALL CONTROL
Select
Register
5.
"Set MAX_RSP_SEQ to 1." on page 191
Screen: CDMA CELL SITE CONFIGURATION
Enter value in:
Max Rsp Seq
6.
"Measure access probe power." on page 192
.
Press: CALL key
Screen: CDMA CALL CONTROL
Observe:
Page Sent
,
Access Probe
,
Connected
7.
"Change the access parameters message." on page 193
Screen: CDMA CELL SITE CONFIGURATION
Enter value in:
Nom Power
,
Init Power
,
Power Step
,
Num Step
,
Max Rsp Seq
8.
"Measure access probe power." on page 194
.
Press: CALL key
Screen: CDMA CALL CONTROL
Observe:
Page Sent
,
Access Probe
,
Connected