![HP HP 8924C Application Manual Download Page 192](http://html.mh-extra.com/html/hp/hp-8924c/hp-8924c_application-manual_166201192.webp)
192
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amtxtst.fb
Chapter 4, CDMA Transmitter Tests
Measuring Access Probe Output Power
6. Measure access probe power.
HP-IB Syntax
"CDMA:CALL:MAKE" !pages the MSUT
Manual Operation:
1. Press the CALL key on the Test Set’s front
panel to page the MSUT.
Nominal settings for the Spectrum Analyzer are
shown.
The figure above shows one access probe sequence, consisting of five access probes.
After the CALL key is pressed on the Test Set, the Page Sent annunciator will light to indicate that a paging
message was sent to the MSUT.
The Access Probe annunciator will light to indicate that the mobile station has transmitted at least one access
probe sequence in an attempt to gain system access.
Spectrum Analyzer:
Sweep Time: 5 s
Ref: -15.0 dBm
1 dB/Div
Span: 0 Hz