Chapter 4, CDMA Transmitter Tests
Measuring Access Probe Output Power
2. Set the test parameter Î
or
.
HP-IB Syntax
"DISP CCNT" !displays the CDMA CALL CONTROL screen.
"CDMA:CELL:ASEC -75" !sets I
or
to -75 dBm/BW.
Manual Operation:
1. Press the CALL CTRL key to display the
CDMA CALL CONTROL screen.
2. Position the cursor at the Sector A Power field.
3. Set the value with the DATA keys. Refer to the
TIA/EIA IS-98 performance standards for test
parameters.
4. Press the ENTER key or the knob to enter the
value.
The value for Sector A Power is referred to as I
or
.
RF Channel
To Screen
CDMA CALL CONTROL
MS FER Report
384
Call Status
Softer
Handoff
Begin/End
MS FER
Report
Interval
5 frames
by # frames
On/Off
Answer Mode
Auto/Manual
Sector A
Power
-75.0
dBm/BW
dBm
%
----
Transmitting
Page Sent
Access Probe
Connected
Softer Handoff
Hard Handoff
Avg Power
Call Limit
None/Page
Register
by # frames
On/Off
5
Traffic
Data Mode
Svc Opt 1
Data Type
Echo
Echo Delay
Short
Avg Power
Power Zero
MS ID
Phone Num
5099214001
MS Database
ESN
1234
CDMA
CALL CNTL
Analog
RX TEST
Cnfig
PRNT CNFG