Chapter 9, 264 Using the Analog Call Processing Subsystem
Using Manual (Front-Panel) Control
12. Position the cursor next to the
Active
field and select it. The
Active
annunciator
will light when the control channel is turned on.
NOTE:
If the
Cntl Channel
field or the
System Type
field were modified in step 11.a. or b.
the control channel will already be active since modifying these fields automatically
activates the control channel.
13. Position the cursor next to the
Register
field and select it. The
Register
annunciator will light while the registration is in process. If the mobile station responds
properly on the reverse control channel the message
RECC Return
will flash momen-
tarily in the upper right-hand corner of the CALL CONTROL screen and the mobile
station phone number, ESN and decoded SCM will be displayed in the upper right-hand
portion of the CALL CONTROL screen. The information in the
MS Id
field will be
updated. The
Active
annunciator will light when the registration has successfully
completed.
Position the cursor next to the
Page
field and select it. The
Page
annunciator will
light while the simulated base station pages the mobile on the forward control channel.
If the mobile station responds properly on the reverse control channel the message
RECC Return
will flash momentarily in the upper right-hand corner of the CALL
CONTROL screen. The
Access
annunciator will then light while the simulated base
station sends the mobile station an alert order on the assigned voice channel. The
mobile station should ring. Press the SEND key on the mobile station. The
Connect
annunciator will light if the mobile station properly signals the simulated
base station when the SEND key is pressed. The mobile station is now connected to
the simulated base station on the assigned voice channel. The left-hand subfields in the
Voice Channel Assignment
section will be updated with the voice channel
assignment information (that is - the “
-
” will be replaced with appropriate informa-
tion).
14. Position the cursor on the
Display
field and select the
Meas
display. The upper right-
hand portion of the CALL CONTROL screen will display modulation quality measure-
ments of the mobile station’s RF carrier. The Test Set’s data functions, such as the av-
erage (
AVG
) function, can be used with any of the measurements. Measurement units
can also be changed as desired.
15. Position the cursor on the
Order
field and select it. A
Choices:
menu is displayed
showing the various power levels which the mobile station can be set to. Position the
cursor next to the desired power level and select it. The simulated base station will then
signal the mobile station with an order to set it’s power level. If the mobile responds
properly the measured value in the
TX Power
field will change.
16. Position the cursor on the
Display
field and select the
Data
display.