27
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amcal.fb
Chapter 1, Calibrating the Test Set
Calibrating CDMA Channel Levels
Cha
p
te
r 1
Calib
ra
ting
The
Tes
t Set
1. Load the PCB_CAL procedure.
HP-IB Syntax
"DISP TEST" ! displays the TESTS (Main Menu) screen.
"TEST:PROC:LOC ’ROM’" ! selects ROM as the test procedure location.
"TEST:PROC:NAME ’PCB_CAL’" selects the file named "PCB_CAL"
Manual Operation:
1. Press the TESTS key.
2. Select ROM from the list of choices for the
Select Procedure Location field.
3. Select PCB_CAL from the list of choices for
the Select Procedure Filename field.
The TESTS (Main Menu) screen provides access to the Test Set’s internal IBASIC controller. You can load,
run, and customize procedures on this screen.
TESTS (Main Menu)
ROM
Select Procedure Location:
PCB_CAL
Select Procedure Filename: