Chapter 3, CDMA Receiver Tests
Measuring Receiver Sensitivity and Dynamic Range
1. Make a Service Option 2 call.
See
"Setting up a Call" on page 52
if you are not familiar with how this procedure
is performed with the Test Set.
After setting up the call, return to this procedure.
NOTE:
The RF path loss between the RF output of the Test Set and the RF input of the MSUT must
be corrected for. Correcting for path loss by entering an RF level offset allows the Test Set
analyzer’s auto-ranging algorithm is to work correctly throughout the operating range of
the MSUT. Refer to