92
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amrcvrt.fb
Chapter 3, CDMA Receiver Tests
Measuring Receiver Sensitivity and Dynamic Range
5. Set the data rate.
HP-IB Syntax
"MEAS:CDMA:CALL:TRAF:DATA:RATE ‘FULL’"
Manual Operation:
1. Position the cursor at the Data Rate field.
2. Press the knob to display a list of choices.
3. Position the cursor at the desired data rate.
Refer to the TIA IS-98 performance standards
for recommended data rate specifications.
4. Press the knob to set the data rate.
Service Option 2 data rate choices: Full=9600 bps, Half=4800 bps, Quarter=2400 bps, Eighth=1200 bps
Service Option 9 data rate choices: Full=14400 bps, Half=7200 bps, Quarter=3600 bps, Eighth=1800 bps
To Screen
CDMA CELLULAR MOBILE RECEIVER TEST
FER
Test Status
Display
Interim
Results
Yes/No
Max Frames
On/Off
Sector A
Power
-105.0
dBm/BW
%
Connected
Svc Opt 2
Testing
Passed
Failed
Max Frames
Errors Counted
Traffic
Data Mode
Svc Opt 2
Data Rate
Full
Eb/Nt
CDMA
CALL CNTL
Analog
RX TEST
Cnfig
PRNT CNFG
Frames Counted
Meas Cntl
Single/Cont
Arm
Disarm
Confidence
95.00
FER Specs
0.50
----
Traffic
-15.6
dB
AWGN
Off
%
%