Chapter 1, Calibrating the Test Set
Determining RF Path Loss
Cha
p
te
r 1
Calib
ra
ting
The
Tes
t Set
3. Determine a reference for the path loss measurement.
Manual Operation:
1. Press the CDMA SCRNS - TX TEST key to
display the CDMA CELLULAR MOBILE
TRANSMITTER TEST screen.
2. Position the cursor at the Sector A Power field.
3. Set the value to -11.0 dBm/BW with the DATA
keys.
4. Position the cursor at the units-of-measure field
and press the knob.
5. Press and release the SHIFT key, then press the
INCR
÷
10 key to set a 0 dBm reference.
CDMA CELLULAR MOBILE TRANSMITTER TEST
-11.0
Sector A
Power
dB/BW
Avg Power
dBm