Chapter 1, Calibrating the Test Set
Correcting for RF Path Loss
1. Enter the path loss from the Test Set to the MSUT.
If you do not know the path loss for your connecting hardware,
Manual Operation:
1. Press and release the SHIFT key and then press
the TESTS key to display the CONFIGURE
screen.
2. Position the cursor in front of the appropriate
field or fields below the RF Level Offset field
and enter a value for RF path loss.
CONFIGURE
RF In/Out
- X .X
dB
Example: If the measured loss is 2 dB, and you are using the RF In/Out port, enter -2 dB in the RF In/Out
field. When the RF Level Offset is turned on, the displayed Average or Channel Power measurement will be
increased by 2 dB and the displayed Sector A, Sector B, AWGN, and RF Power outputs will be decreased by
2 dB. No actual level changes occur as a result of turning on RF Level Offset.
Duplex Out
- X .X
Antenna In
- X .X