Leveling Mode
Normal
Function Group
ALC
Menu Map
Description
This
lets you set the leveling mode of the synthesizer to
continuous leveling at the desired leveling point. In this mode, the
RF OUTPUT is controlled by the automatic level control (ALC)
circuit, otherwise referred to as the leveling loop. The attenuator
works in conjunction with the ALC to achieve the full range of
power levels. At factory preset, ALC normal is the default state. An
asterisk next to the key label indicates that this feature is active.
Programming Codes
SCPI: POWer:ALC:STATe
Analyzer:
Al, internal normal;
external normal;
external
power meter normal;
source module normal.
See Also
Leveling Mode
Search
Function Group
ALC
Menu Map
Description
This
causes the ALC to switch off once the desired power
level is reached. When this leveling mode is activated, or when
power, or frequency is changed, the synthesizer switches to CW
frequency and closes the ALC system until the desired power level is
reached. The synthesizer reverts to its original frequency/modulation
state and opens the ALC system. This mode is similar to ALC off
mode and is useful for narrow pulse applications. An asterisk next to
the key label indicates that this feature is active.
Programming Codes
SCPI:
POWer:ALC:STATe
Analyzer:
Pulse Modulation
“Working with Spectrum Analyzers,” in Chapter 1.
Operating and Programming Reference
HP 8360
User’s Handbook
Summary of Contents for 8360
Page 97: ......
Page 138: ......
Page 150: ......
Page 318: ......
Page 336: ......
Page 352: ......
Page 360: ......
Page 397: ...2b Menu Maps Menu Maps 2b 1 ...
Page 398: ......
Page 399: ......
Page 400: ......
Page 401: ......
Page 402: ......
Page 403: ......
Page 404: ......
Page 405: ......
Page 406: ......
Page 435: ...Figure 3 3 Removing the Side StraPs and Feet NSTALLATION 3 ...
Page 461: ......
Page 463: ...5 4 Change B HP 8380 User s Handbook ...
Page 467: ...5 8 Change A HP 8360 User s Handbook ...
Page 481: ...optioll9l3RackFhgeKit 5 22 Change A HP 8360 User s Handbook ...