Internal Pulse Generator Period
Internal Pulse Mode Gate
Turns on the internal pulse mode during the
positive cycle of the externally generated pulse.
Internal
Pulse Mode
Trigger
Triggers on the leading edge of the external pulse
input.
Programming Codes
SCPI:
NONE, see the individual
listed.
Analyzer:
NONE
See Also
a so see “Modulation” and “Pulse”.
Internal Pulse Generator Period
Function Group
Menu Map
Description
This
(Option 002 only) lets you set a value for the internal
pulse generator’s pulse period. The pulse is adjustable from
300 ns to 400 ms with ‘25 ns resolution. The factory preset default
is 2 ms pulse period. When this feature is active, its current value is
displayed in the active entry area.
Since period and rate are inversely related, if both are given values,
only the last one will be applied which will cause the first one to be
recalculated. Use the one that is convenient for your application. For
example, if you set the pulse period, do not change the pulse rate
(the synthesizer automatically adjusts the rate to match the period.)
Programming Codes
SCPI:
PULS:INTernal:PERiod <num>[time
Analyzer:
NONE
See Also
a so see “Pulse” and “Modulation”.
HP 8360
User’s Handbook
Operating and Programming Reference
Summary of Contents for 8360
Page 97: ......
Page 138: ......
Page 150: ......
Page 318: ......
Page 336: ......
Page 352: ......
Page 360: ......
Page 397: ...2b Menu Maps Menu Maps 2b 1 ...
Page 398: ......
Page 399: ......
Page 400: ......
Page 401: ......
Page 402: ......
Page 403: ......
Page 404: ......
Page 405: ......
Page 406: ......
Page 435: ...Figure 3 3 Removing the Side StraPs and Feet NSTALLATION 3 ...
Page 461: ......
Page 463: ...5 4 Change B HP 8380 User s Handbook ...
Page 467: ...5 8 Change A HP 8360 User s Handbook ...
Page 481: ...optioll9l3RackFhgeKit 5 22 Change A HP 8360 User s Handbook ...