Doubler Amp Mode On
Function Group
Menu Map
Description
Programming Codes
See Also
POWER
5
This
is applicable to instrument models with a doubler
installed. The doubler has an integral amplifier whose operation is
controlled by the instrument firmware. This
turns off the
automatic mode of operation and turns on the amplifier so that it is
always used. This is an
unspecified mode
of operation since it can
cause increased harmonics and degraded dynamic range at some
frequencies. This
has no effect on instruments without a
doubler.
An asterisk next to the key label indicates that this feature is active.
POWer:AMPLifier:STATE
POWer:AMPLifier:STATE?
Analyzer:
NONE
Dblr
Menu
Dwell Coupled
Function Group
FREQUENCY
Menu Map
Description
This
lets you couple the dwell time for points in the stepped
frequency sweep mode to the ramp sweep mode sweep time. The
equation to determine the dwell time in the dwell coupled mode is as
follows:
Coupled Dwell Time (sweep time) (number of step points)
An asterisk next to the key label indicates that this feature is active.
Programming Codes
SCPI:
SWEep[:FREQ
Analyzer:
NONE
See Also
Step Sup
Menu
Operating and Programming Reference
HP 8360
User’s Handbook
Summary of Contents for 8360
Page 97: ......
Page 138: ......
Page 150: ......
Page 318: ......
Page 336: ......
Page 352: ......
Page 360: ......
Page 397: ...2b Menu Maps Menu Maps 2b 1 ...
Page 398: ......
Page 399: ......
Page 400: ......
Page 401: ......
Page 402: ......
Page 403: ......
Page 404: ......
Page 405: ......
Page 406: ......
Page 435: ...Figure 3 3 Removing the Side StraPs and Feet NSTALLATION 3 ...
Page 461: ......
Page 463: ...5 4 Change B HP 8380 User s Handbook ...
Page 467: ...5 8 Change A HP 8360 User s Handbook ...
Page 481: ...optioll9l3RackFhgeKit 5 22 Change A HP 8360 User s Handbook ...