Inside the Sequence Operation State.
Figure
illustrates the
operation of the sequence operation state.
The downward entrance to the Sequence Operation State signals
that some instrument dependent action should begin at once. An
upward exit is not allowed until the instrument signals that its
action is complete. Note that complete can be defined differently for
different instruments. For example, consider an instrument that can
sweep a range of frequencies starting with and ending with
The
action-complete signal can be defined to coincide with the output of
either or
Figure
Inside the Sequence Operation State
Common Trigger
In the previous paragraphs, you learned about the basic building
Configurations
blocks allowed in a SCPI trigger system. Generally, an instrument
implements only a portion of the trigger features available. These
paragraphs discuss the simplest configurations:
and TRIG.
The
Configuration
The
configuration is the simplest possible trigger configuration.
It uses no event detection states, and requires only two subsystems
for programming,
and
.
All SCPI instruments
implement these two subsystems
Getting Started Programming
Summary of Contents for 8360
Page 97: ......
Page 138: ......
Page 150: ......
Page 318: ......
Page 336: ......
Page 352: ......
Page 360: ......
Page 397: ...2b Menu Maps Menu Maps 2b 1 ...
Page 398: ......
Page 399: ......
Page 400: ......
Page 401: ......
Page 402: ......
Page 403: ......
Page 404: ......
Page 405: ......
Page 406: ......
Page 435: ...Figure 3 3 Removing the Side StraPs and Feet NSTALLATION 3 ...
Page 461: ......
Page 463: ...5 4 Change B HP 8380 User s Handbook ...
Page 467: ...5 8 Change A HP 8360 User s Handbook ...
Page 481: ...optioll9l3RackFhgeKit 5 22 Change A HP 8360 User s Handbook ...