Installation
Requirements
V
entilation
Requirements
The
HP
4155A/4156A
has
one
cooling
fan,
and
HP
41501A
has
two
cooling
fans
.
T
o
ensure
adequate
airow
,
make
sure
that
there
is
adequate
clearance
around
the
cooling
fans:
6
inches
(150
mm)
behind,
3
inches
(70
mm)
sides
,
and
0.5
inch
(12
mm)
above
and
below
.
If
the
airow
is
restricted,
the
internal
operating
temperature
will
be
higher
,
reducing
the
instrument's
reliability
or
causing
the
instrument's
thermal-protection
circuits
to
automatically
switch
o
the
instrument.
Operating
Environment
The
HP
4155A/4156A
and
HP
41501A
must
be
operated
within
the
following
environmental
conditions:
T
emperature:
5
C
to
40
C
(41
F
to
104
F)
Humidity:
15%
to
80%
RH
at
40
C
(104
F)
Cleaning
T
o
prevent
electrical
shock,
never
use
a
wet
cloth
when
cleaning
the
HP
4155A/4156A.
Always
use
a
slightly
damp
or
dry
cloth
to
clean
HP
4155A/4156A.
2-7
Summary of Contents for 4155A
Page 2: ...Semiconductor Parameter Analyzer User s Task Guide ...
Page 6: ......
Page 7: ......
Page 8: ......
Page 9: ......
Page 19: ...Contents ...
Page 20: ...1 Introducing the HP 4155A 4156A ...
Page 47: ...Introducing the HP 4155A 4156A An Overview of Functions ...
Page 48: ...2 Installation ...
Page 84: ...Installation Installing Accessories Attaching HP 16441A R Box to Shielding Box 2 37 ...
Page 94: ...3 Making a Measurement ...
Page 161: ...Making a Measurement Stress Force ...
Page 162: ...4 Analyzing Measurement Results ...
Page 206: ...5 Filer ...
Page 211: ...Filer File Operations Example The following gure shows example that displays le catalog 5 6 ...
Page 233: ...Filer File Operations ...
Page 234: ...6 If Y ou Have A Problem ...
Page 276: ...7 Manual Changes Depending on ROM Version ...
Page 279: ...Manual Changes Depending on ROM Version Change 1 ...