If
Y
ou
Have
A
Problem
If
Errors
Occur
1xx08
AD
converter
test
reached
timeout.
Measurement
unit
failed
AD
converter
test
as
a
pretest
for
calibration
or
self-test.
1xx11
Overvoltage
occurred
for
a
measurement
unit.
1xx12
Overcurrent
occurred
for
a
measurement
unit.
1xx15
Measurement
units
that
are
not
supported
are
detected.
1xx19
Emergency
occurred
but
the
cause
is
unknown.
This
is
displayed,
for
example
,
when
unit
is
known
but
cause
is
unknown.
1xx90
AD
converter
test
reached
timeout
during
calibration
or
self-test
for
a
measurement
unit.
1xx91
FIFO
(rst-in,
rst-out)
for
AD
converter
overowed
because
SMU
controller
takes
long
time
to
read
measurement
data.
1xx92
Calibration
or
diagnostics
was
aborted
by
an
emergency
or
*RST
command.
1xx94
HP
4155A/4156A
was
turned
on
before
HP
41501A.
1xx97
Communication
failed
between
HOST
controller
and
SMU
controller
.
Or
calibration/diagnostics
was
performed,
but
HOST
controller
couldn't
receive
the
result
from
SMU
controller
.
10030
V
SUs
and
VMUs
failed
default
test
of
calibration.
10031
V
SUs
and
VMUs
failed
function
check.
10032
V
SUs
failed
gain
or
oset
calibration.
10033
VMUs
failed
gain
or
oset
calibration.
6-18
Summary of Contents for 4155A
Page 2: ...Semiconductor Parameter Analyzer User s Task Guide ...
Page 6: ......
Page 7: ......
Page 8: ......
Page 9: ......
Page 19: ...Contents ...
Page 20: ...1 Introducing the HP 4155A 4156A ...
Page 47: ...Introducing the HP 4155A 4156A An Overview of Functions ...
Page 48: ...2 Installation ...
Page 84: ...Installation Installing Accessories Attaching HP 16441A R Box to Shielding Box 2 37 ...
Page 94: ...3 Making a Measurement ...
Page 161: ...Making a Measurement Stress Force ...
Page 162: ...4 Analyzing Measurement Results ...
Page 206: ...5 Filer ...
Page 211: ...Filer File Operations Example The following gure shows example that displays le catalog 5 6 ...
Page 233: ...Filer File Operations ...
Page 234: ...6 If Y ou Have A Problem ...
Page 276: ...7 Manual Changes Depending on ROM Version ...
Page 279: ...Manual Changes Depending on ROM Version Change 1 ...