Modulating Source
The function generator will accept an internal modulating signal,
an external modulating signal, or both.
•
Modulating source: Internal-External (both) or External only.
The default is Both (internal-external).
•
The External modulating source is always enabled.
•
When both sources are enabled (
internal-external
), the function
generator adds the internal and external modulating signals
(the carrier waveform is actually modulated with two waveforms).
•
When the internal source is disabled (
external only
), the carrier
waveform is modulated with the external waveform. The
Ext
annunciator turns on to indicate that the function generator is
expecting a modulating signal on the rear-panel
AM Modulation
terminal.
•
You apply the external modulating waveform to the
AM Modulation
terminal. The
modulation depth
is controlled by the signal level
present (5 volts
peak
corresponds to 100% modulation).
•
Front-Panel Operation
:
2: AM SOURCE (MOD MENU)
•
Remote Interface Operation
:
AM:SOURce {BOTH|EXTernal}
5 V
0 V
-5 V
3
Chapter 3 Features and Functions
Amplitude Modulation (AM)
75
Summary of Contents for 33120A
Page 6: ...1 Quick Start 1 ...
Page 20: ...2 Front Panel Menu Operation 2 ...
Page 44: ...3 Features and Functions 3 ...
Page 115: ...8 Specifications 8 ...