Possible Conflict with Output Frequency
: The duty cycle is auto-
matically adjusted if you select a frequency that is not valid with the
present duty cycle. For example, if you set the duty cycle to 70% and
then change the frequency to 8 MHz, the function generator will
automatically adjust the duty cycle to 60% (the upper limit for this
frequency).
From the front panel, “
%
DUTY LIMIT
” is displayed and the duty cycle is
adjusted. From the remote interface, a
-221, “Settings conflict”
error
is generated and the duty cycle is adjusted.
•
Front-Panel Operation
: To set the duty cycle, press
Shift % Duty
(the value appears on the display for approximately 10 seconds).
Then use the knob, arrow keys, or “Enter Number” mode to set the
duty cycle.
•
Remote Interface Operation
:
PULSe:DCYCle {<
percent
>|MINimum|MAXimum}
The
APPLy
command automatically sets the duty cycle to 50% for
square waves.
3
Chapter 3 Features and Functions
Output Configuration
67