Burst Rate
The
burst rate
defines the frequency at which
internally
triggered bursts
are generated. The burst rate frequency defines the interval between
bursts.
Used only in the internal triggered burst mode.
Keep in mind that the burst rate is different than the “carrier frequency”
which specifies the frequency of the bursted signal.
•
Burst rate: 10 mHz to 50 kHz.
The default is 100 Hz.
•
The burst rate setting is used only when
internal
triggering is
enabled (triggered mode). The burst rate is ignored when single
triggering or external triggering is enabled.
•
When the
single
or
external
trigger source is selected (triggered mode),
the
burst count
and
burst phase
remain in effect but the
burst rate
is ignored.
•
When the
external gate
burst source is selected (“gated” burst mode),
the burst count, burst rate, burst phase, and burst trigger source are
ignored (these parameters are used for the triggered burst mode only).
•
It is possible to specify a burst rate which is too fast for the function
generator to output with the specified carrier frequency and burst
count. If the burst rate is too high, the function generator will
internally adjust it as needed to continuously re-trigger the burst.
The adjustment is handled internally by the function generator
(the burst rate displayed or queried will be the same as specified).
•
Front-Panel Operation
:
5: BURST RATE (MOD MENU)
•
Remote Interface Operation
:
BM:INTernal:RATE {<
frequency
>|MINimum|MAXimum}
Chapter 3 Features and Functions
Burst Modulation
88