Baud Rate Selection (
RS
-232)
You can select one of six baud rates for
RS
-232 operation. The rate is set
to
9600 baud
when the function generator is shipped from the factory.
The baud rate can be set from the front-panel only.
•
Select one of the following: 300, 600, 1200, 2400, 4800, or
9600
baud.
The factory setting is 9600 baud
.
•
The baud rate selection is stored in
non-volatile
memory, and
does not
change when power has been off or after a remote interface reset.
•
Front-Panel Operation
:
3: BAUD RATE (I/O MENU)
See also “To set the baud rate,” on page 219.
Parity Selection (
RS
-232)
You can select the parity for
RS
-232 operation. The function generator
is configured for
no parity and 8 data bits
when shipped from the factory.
The parity can be set from the front-panel only.
•
Select one of the following:
None
(
8 data bits
), Even (
7 data bits
),
or Odd (
7 data bits
). When you set the parity, you are indirectly
setting the number of data bits.
•
The parity selection is stored in
non-volatile
memory, and
does not
change when power has been off or after a remote interface reset.
•
To download binary data for arbitrary waveforms over the
RS
-232
interface, you must select
8 data bits
with
no parity
.
•
Front-Panel Operation
:
4: PARITY (I/O MENU)
See also “To set the parity,” on page 220.
Chapter 3 Features and Functions
Remote Interface Configuration
116
Summary of Contents for 33120A