Carrier Frequency
(continued)
•
For
all waveforms
used with burst, if the carrier frequency is set
less than or equal to
100 Hz
, the following relationship applies.
Burst Count
Carrier Frequency
≤
500
seconds
For Carrier
≤
100 Hz
If you attempt to set the carrier frequency to a value that is not valid,
the function generator will automatically adjust the frequency to the
minimum value allowed with the present burst count.
From the front panel, “
MIN VALUE
” is displayed and the frequency is
adjusted. From the remote interface, a
-221, “Settings conflict”
error
is generated and the frequency is adjusted.
•
For the
triggered
burst mode, the sync signal is a
TTL
“low” while the
specified number of cycles is output (for the duration of the burst).
After the specified number of cycles has been output, the sync signal
goes “high” until the next burst. The sync signal is output from the
front-panel
SYNC
terminal.
•
For the
external gated
burst mode, the sync signal is a
TTL
“high”
when the output is positive, relative to zero volts (or the dc offset value).
The signal is a
TTL
“low” when the output is negative, relative to
zero volts (or the dc offset value). The sync signal is output from the
front-panel
SYNC
terminal.
•
Front-Panel Operation
: Press the
Freq
key and select any valid
frequency for the selected function (the present amplitude and offset
voltage are used).
•
Remote Interface Operation
:
FREQuency {<
frequency
>|MINimum|MAXimum}
You can also use the
APPLy
command to select the function,
frequency, amplitude, and offset of the carrier with a single command.
Chapter 3 Features and Functions
Burst Modulation
86