Calibration Message
You can use the calibration message feature to record calibration
information about your function generator. For example, you can store
such information as the last calibration date, the next calibration due
date, the instrument serial number, or even the name and phone
number of the person to contact for a new calibration.
You can record information in the calibration message only from the
remote interface. However, you can read the message from either the
front-panel menu or the remote interface.
•
The calibration message may contain up to 40 characters. However,
the function generator can display only 11 characters of the message
on the front panel (additional characters are truncated).
•
The calibration message is stored in
non-volatile
memory, and
does not
change when power has been off or after a remote
interface reset.
•
Front-Panel Operation
:
4: MESSAGE (CAL MENU)
Read the cal message
•
Remote Interface Operation
:
CALibration:STRing <
quoted string
>
Store the cal message
The following command string shows how to store a calibration message.
"CAL:STR ’CAL 12-10-93’"
Chapter 3 Features and Functions
Calibration Overview
122