Quanta-
φ
rev. C (23 Apr 2010)
Troubleshooting
6-3
Possible contamination of inner
sphere
Internal contamination voids the
sphere warranty.
The sphere inte-
rior may need to be replaced be-
cause the porous Spectralon
®
is
very difficult to decontaminate.
CCD signal > 65 000
counts per accumula-
tion cycle for a blank
excitation signal (> 10
6
cps).
Excessive count rate (CPS) or inte-
gration time
CCD count-rate for the blank exci-
tation band should be adjusted to
between 500 000 and 1 million
counts per second (i.e., 50 000
counts between 0.05 and 0.1 s).
Bandpass too wide
Bandpass should be set for com-
plete resolution of the spectral fea-
tures of the excitation bands and
fluorescence signal areas. Usually
from 1 to 5 nm is sufficient.
Insufficient neutral-density attenua-
tion
Use neutral attenuation on excita-
tion side if the suggested count rate
(500 000–1 000 000 cps) is ex-
ceeded when the bandpass is be-
tween 1 and 5 nm.
PMT signal > 1 million
cps (rate) for blank
excitation signal
Insufficient neutral-density attenua-
tion
Neutral attenuation should be used
on excitation if the suggested count
rate (500 000–1 000 000 cps) is
exceeded when the bandpass is be-
tween 1 and 5 nm.
Bandpass is too wide
Bandpass should be set for com-
plete resolution of the spectral fea-
tures of the excitation bands and
fluorescence signal areas. Usually
from 1 to 5 nm is sufficient.
Summary of Contents for Quanta-phi F-3029
Page 1: ......
Page 2: ......
Page 6: ...Quanta rev C 23 Apr 2010 iv...
Page 28: ...Quanta rev C 23 Apr 2010 Theory of Operation 1 14 Flowchart of method for data acquisition...
Page 30: ...Quanta rev C 23 Apr 2010 Theory of Operation 1 16...
Page 34: ...Quanta rev C 23 Apr 2010 Requirements Unpacking 2 4...
Page 64: ...Integrating Sphere v 1 2 3 Jul 2008 Index 3 30...
Page 66: ...Quanta rev C 23 Apr 2010 Maintenance 4 2...
Page 84: ......
Page 98: ......