Quanta-
φ
rev. C (23 Apr 2010)
Troubleshooting
5-6
Caution:
Do not touch the surfaces of the Integrating
Sphere’s sample-holder or the reflector accessory. This
can degrade its highly reflective surface. Wear clean pro-
tective gloves when handling the interior of the sphere and
reflector.
l
Insert the uncalibrated reflector into
the cuvette sample-holder. Be sure
the square top and bottom of the
reflector are seated properly into the
square depressions in the sample
holder.
2
Scan the reflectance spectrum in the integrating
sphere.
a
Insert the sample holder into
the integrating sphere.
Be sure the plane of the ref-
lector forms a 45° angle be-
tween the excitation and
emission ports.
b
In the FluorEssence™
Experiment Setup
window,
scan the reflectance
spectrum as an emission
scan using the same
experimental conditions as
in Step
i
above.
Excitation
port
Emis-
sion
port