Installation & Use
3-27
5
Analyze the scan for the counts in the Rayleigh
peak.
If the peak is ~10
6
counts, this is acceptable.
6
Rerun the scan if the peak is not ~10
6
counts.
a
Click the
Previous Experiment
button
.
b
Click the
Detectors
icon to see the detectors page.
c
Set the units to
Counts
.
d
Set the
Integration Time
to yield 10
6
counts in the Rayleigh peak.
7
Click the Run button to start the scan.
The
Intermediate Display
appears:
The
Project Name
window may appear,
then the data appear. If the peak is ~10
6
counts, this is acceptable.
8
Remove the blank, and place the sample inside
the integrating sphere.