Note:
See the section later in this chapter if using a photomul-
tiplier tube as the detector.
Continue here if your detector is a CCD.
7
Adjust the Rayleigh excitation peak for the blank
(solvent or solid) to the desired count rate (sug-
gested value
≈ 10
6
counts/s).
We are trying to maximize but not saturate the excitation Rayleigh peak, to ob-
tain an optimal luminescence signal-to-noise ratio.
a
Place the solvent blank cuvette (or blank solid material) into the Inte-
grating Sphere.
b
In the FluorEssence™ toolbar, click the
Experiment Menu
button
.
The
Fluorescence Main Experi-
ment Menu
appears:
c
Click the
Spectra
button.
The
Experiment Type
window ap-
pears.
d
Choose the
Emission
subtype of ex-
periment, and click the
Next >>
but-
ton.
The
Experiment Setup
window ap-
pears: