Quanta-
φ
rev. C (23 Apr 2010)
Installation & Use
3-12
h
Enable the emission detector, the
Correction
checkbox, and the
Dark
Offset
checkbox. Set the
Integration Time
to 0.1 s.
i
Add >>
the corrected signal (with
c
subscript) to the
Formulas
table.
j
Give this file a
Data Identifier
.
k
Choose
1
Accumulations
.
If this is too noisy, you will increase the accumulations later.
l
Click the
Run
button.
The
Interme-
diate Display
appears.
m
A
Warning
win-
dow may appear
that the
length range
cludes values
outside of the
correction-factor
file’s range.
Summary of Contents for Quanta-phi F-3029
Page 1: ......
Page 2: ......
Page 6: ...Quanta rev C 23 Apr 2010 iv...
Page 28: ...Quanta rev C 23 Apr 2010 Theory of Operation 1 14 Flowchart of method for data acquisition...
Page 30: ...Quanta rev C 23 Apr 2010 Theory of Operation 1 16...
Page 34: ...Quanta rev C 23 Apr 2010 Requirements Unpacking 2 4...
Page 64: ...Integrating Sphere v 1 2 3 Jul 2008 Index 3 30...
Page 66: ...Quanta rev C 23 Apr 2010 Maintenance 4 2...
Page 84: ......
Page 98: ......