XNX Universal Transmitter
Section 3 - Calibration
94
3.2.4 Zero and Span Calibration of XNX EC Hydrogen
Sulfide (
H
2
S) Sensors
Caution:
Before initial calibration, allow the sensor to stabilize for 30 minutes
after applying power. When in zero and span calibration modes, the current
output from the sensor is inhibited (default 2mA) to avoid false alarms.
Caution:
Recalibrate if the temperature of local environment has varied by
more than ±15°C from the temperature of calibration.
Hydrogen Sulfide sensors can be affected by extreme humidity
changes� A sudden increase in ambient humidity can result in a
short-term positive drift in the instrument’s reading� A sudden
decrease in ambient humidity can result in a short-term negative
drift in the instrument’s reading� These are most likely to be
noticed during calibration with dry or cylinder gas�
When calibrating hydrogen sulfide cartridges, the following
should be taken into account while following the procedure in
Section 3�2�2
�
1� To zero the sensor, use a compressed gas cylinder of
20�9%Vol oxygen (not nitrogen)� Do not use background
air�
If a span calibration is to be performed, the span calibration gas
should be applied to the sensor immediately after the zeroing
procedure� Do not allow the sensor to return to ambient air
conditions�
Warning:
Long-term exposure (> 20 minutes) to concentrations exceediing the
full-scale range of the sensor can cause it to lose sensitivity. The ouput of the
sensor may then decrease in value even though high levels of toxic gas are still
present. Before re-calibrating the transmitter, verify the absence of gas.
3.2.5 705/705HT Calibrating
For complete calibration and configuration information, see the
Type 705 Operating Instructions (P/N: 00705M5002)�
3.2.6 Sensepoint/Sensepoint HT Calibrating
For complete calibration and configuration information, see the
Sieger Sensepoint Technical Handbook (P/N: 2106M0502)�
3.2.7 Calibrating the Searchpoint Optima Plus
Complete calibration and configuration information can be
found in the Searchpoint Optima Plus Operating Instructions
(P/N:2108M0501)� If properly installed and maintained, the
Searchpoint Optima Plus sensor will not require routine
calibration� This is due to the inherent stability of the IR
absorption process and the unit’s fully compensated optical
configuration�
1� From the Calibration menu, select the Gas Calibration option�
Figure 204. Calibration menu
Summary of Contents for XNX
Page 1: ...Technical Manual XNXTM Universal Transmitter ...
Page 2: ......
Page 6: ...XNX Universal Transmitter Table of Contents 4 Table of Contents ...
Page 11: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 9 1 Introduction ...
Page 91: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 89 3 Calibration ...
Page 104: ...XNX Universal Transmitter Section 3 Calibration 102 ...
Page 105: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 103 4 Maintenance ...
Page 128: ...XNX Universal Transmitter Section 5 Warnings Faults 126 ...
Page 129: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 127 6 Specifications ...
Page 152: ...XNX Universal Transmitter Section 6 Specifications 150 Figure 232 XNX BT Configuration ...
Page 161: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 159 7 Control Drawings ...
Page 162: ...XNX Universal Transmitter Section 7 Control Drawings 160 7 1 XNX UL INMETRO ...
Page 163: ...XNX Universal Transmitter Section 7 Control Drawings 161 ...
Page 164: ...XNX Universal Transmitter Section 7 Control Drawings 162 ...
Page 165: ...XNX Universal Transmitter Section 7 Control Drawings 163 7 2 XNX UL CSA FM ...
Page 166: ...XNX Universal Transmitter Section 7 Control Drawings 164 ...
Page 167: ...XNX Universal Transmitter Section 7 Control Drawings 165 ...
Page 192: ...XNX Universal Transmitter Appendix B Modbus Protocol 190 ...
Page 193: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 191 Appendix C Warranty ...
Page 195: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 193 Index ...
Page 199: ......