XNX Universal Transmitter
Section 6 - Specifications
142
6.2.6 XNX MPD Sensor Performance Data
Sensor
Type
Gas
Typical
Response
Time
(T50) sec
Typical
Response
Time
(T90) sec
Maximum
Range
Accuracy
(% of full scale or
% of applied gas)
Drift
Over
Time
Operating
Humidity
Operating Temperature
Operating
Pressure
Operating
Air Speed
Min
Max
MPD-IC1
Carbon Dioxide
<30
<70
5.00 %Vol
±5%FS or ±15%
< 3%/yr
0-95%
RH non-
condensing
-20°C / -4°F +50°C / 122°F 80kPa ~ 110kPa 0 ~ 6m/sec
MPD-IV1
Methane
<15
<30
5.00 %Vol
±5%FS or ±15%
< 3%/yr
-20°C / -4°F +50°C / 122°F 80kPa ~ 110kPa 0 ~ 6m/sec
MPD-IF1
Propane
<15
<30
100 %LEL
±5%FS or ±15%
< 3%/yr
-20°C / -4°F +50°C / 122°F 80kPa ~ 110kPa 0 ~ 6m/sec
MPD-CB1
Propane
<15
<30
100 %LEL
±5%FS or ±15%
< 3%/yr
-40°C /
-40°F
+65°C / 149°F 80kPa ~ 120kPa 0 ~ 6m/sec
Methane
<10
<30
Hydrogen
<10
<30
Butane-2
<15
<40
Nonane
<20
<50
NOTES:
• Response times may vary depending upon molecular weight, size, and structure�
• CSA approved hydrogen sensors are MPDUT-CB1 and 705 STD�
• DEKRA EXAM approved sensors are MPDAM CB1 and SPHT
• Data taken at 20-25°C� Contact Honeywell Analytics for additional data or details�
• Response times may increase at lower temperatures�
• Data represents typical values without optional accessories attached�
• System conditioning may be required to achieve stated results� Contact Honeywell Analytics for details�
• Performance figures are measured using a sample humidity of 50% RH�
• Performance figures are measured between 40 and 60% of full scale�
• Performance figures are measured by test units calibrated at 50% of full scale�
• Use of the weatherproof cap will increase response times�
• FM 6340 performance approval based on MPD-IC1 with SPXCDWP T50<60 T90<150�
• Use of the weatherproof cap will increase response times�
• FM 6340 performance approval based on MPD-IC1 with SPXCDWP T50<60 T90<150�
Summary of Contents for XNX
Page 1: ...Technical Manual XNXTM Universal Transmitter ...
Page 2: ......
Page 6: ...XNX Universal Transmitter Table of Contents 4 Table of Contents ...
Page 11: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 9 1 Introduction ...
Page 91: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 89 3 Calibration ...
Page 104: ...XNX Universal Transmitter Section 3 Calibration 102 ...
Page 105: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 103 4 Maintenance ...
Page 128: ...XNX Universal Transmitter Section 5 Warnings Faults 126 ...
Page 129: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 127 6 Specifications ...
Page 152: ...XNX Universal Transmitter Section 6 Specifications 150 Figure 232 XNX BT Configuration ...
Page 161: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 159 7 Control Drawings ...
Page 162: ...XNX Universal Transmitter Section 7 Control Drawings 160 7 1 XNX UL INMETRO ...
Page 163: ...XNX Universal Transmitter Section 7 Control Drawings 161 ...
Page 164: ...XNX Universal Transmitter Section 7 Control Drawings 162 ...
Page 165: ...XNX Universal Transmitter Section 7 Control Drawings 163 7 2 XNX UL CSA FM ...
Page 166: ...XNX Universal Transmitter Section 7 Control Drawings 164 ...
Page 167: ...XNX Universal Transmitter Section 7 Control Drawings 165 ...
Page 192: ...XNX Universal Transmitter Appendix B Modbus Protocol 190 ...
Page 193: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 191 Appendix C Warranty ...
Page 195: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 193 Index ...
Page 199: ......