Input
Calibration -
Restoring
Factory Calibration
Release M
DR4500A Classic Series Circular Chart Recorder With or Without Control Product Manual
177
April 2017
Step
Operation
Press
Action/Result
▲
or
▼
Until you see:
Upper Display:
0-10V
Lower Display:
INn TYPE
ATTENTION
If you are already using a 0-10V input,
change to a 0-5V input type.
FUNC
Until the upper display rolls through the input types and
returns to
Upper Display:
0-10V
Lower Display:
INn TYPE
2
(cont)
▲
or
▼
Until you change the input selection in the upper display
back to the proper selection. You will see:
Upper Display:
Original input selection that
matches your type of sensor.
Lower Display:
INn TYPE
To commit press FUNC before pressing LOWR DISP
3
Return to normal
operating mode
LOWR
DISP
To return to Normal operating mode.
The factory calibration will be restored. If the problem is
not corrected, contact the Honeywell Technical
Assistance Center.
1-800-423-9883 USA and Canada
CAUTION
A restored factory calibration overwrites any previous field calibration done for the input. Be
sure to protect any field calibration from accidental overwrites by configuring the appropriate
“LOCKOUT” selection after calibration. See the
Section 3 - Configuration
for specific
instructions to set the lockout.