5.2.5 F
ULL PARAMETER PROGRAMMING
Full Parameter Access is the highest programming level to access all features provided by the NIVOCAP, level, volume and weight can also be
displayed in engineering units.
Description of the parameters can be found under
6. paragraph.
Keys
Function
E
+ (at least for 3 sec)
Enter/quit Full Parameter Programming mode. Entering programming mode parameter modified last time is blinking.
During programming „PROG” is lit while the screen is as below:
yy Parameter address (
P01
,
P02
…
P99
)
xxxx Parameter value (dcba)
Bargraph
Measuring is going on during programming in accordance with the old parameter set. New parameter set will be valid after returning to the
Measurement Mode.
Steps and indications of the Full Parameter Access programming mode
Pressing
Keys
While
Parameter address is blinking
While
Parameter value is blinking
E
Selecting parameter address and step to the parameter value
Modification of the parameter values will be saved.
Return to the parameter address.
+
Cancel all modifications of the actual programming phase.
Pressing for 3 sec is required while CANCEL will be displayed for
warning
Modification of the parameter value will be neglected
Return to the parameter address without saving the
modifications.
+
Device will be returned to Factory Default.
Since all parameters will be overwritten “LOAD” will appear on the
display first:
– to confirm, press
E
– to escape, press any other key
Default parameter will be loaded
(and saved by pressing ENTER
E
).
Move blinking (changeability) of the digit to the left
/
Modify the blinking digit (increase, decrease) or scroll up/down