80
Specifications
Accuracy specifications
Accuracy guarantee
conditions
Accuracy guarantee duration: 1 year
Accuracy guarantee duration after adjustment made by Hioki: 1 year
Accuracy guarantee temperature and humidity range:
23°C ±5°C (73°F ±9°F)
,
80% RH or less (non-condensing)
Number of jaw open/close cycles: 10000 times or less
Input conditions for
guaranteed accuracy
Sine wave input
Effects of external
magnetic fields
DC/AC 60 Hz
,
with a 400 A/m external magnetic field: 0.10 A or less
Effects of conductor
position
At all positions around the jaw’s center-point reference: within ±0.5%
(100 A input
,
f ≤ 100 Hz)
Temperature coefficient
Add “measurement accuracy × 0.1/°C” (excluding 23°C ±5°C [73°F ±9°F]).
Effects of sensor phase
±1° (50 Hz to 60 Hz)
HIOKI CM3286C961-00