Troubleshooting
If damage is suspected
,
check the following before contacting your authorized Hioki distributor
or reseller.
Problem
Cause
Remedy
The instrument
is indicating an
abnormal measured
value.
The measured value is lower
than the lower limit value of
the measuring range.
Wrap the wire around the jaw one or
more times. Wrapping the wire n times
can increase the displayed value by (
n
+
1) times.
The tips of the jaw open.
Close the jaw tips.
The jaw is damaged.
The instrument with its jaw damaged
cannot measure current accurately. Have
the instrument repaired.
Displayed values can frequently fluctuate due to induction potential
even with no input. This
,
however
,
is not a malfunction.