88
Accuracy Table
(4) Balanced 3‑phase/3‑wire power measurement
Automatic range threshold:
Range up: Current RMS value greater than 6000 count
Range down: Current RMS value less than 540 count
Effective
measuring
range
Current RMS
value (
I
RMS
)
0.060 A to 600.0 A
Value must fall within the current measurement range’s guaranteed
accuracy range.
Voltage RMS
value (
U
RMS
)
80.0 V to 600.0 V
Frequency
50 Hz/60 Hz
Balanced
3‑phase/3‑wire
active power
(
P
(3P3W)
)
Balanced
3‑phase/3‑wire
apparent power
(
S
(3P3W)
)
Accuracy
±3.0% rdg ±10 dgt (Power factor =1)
Range
configuration
(minimum
resolution)
Current range
6.000 A
60.00 A
600.0 A
Voltage range 600.0 V
7.200 kW
(0.001 kW)
72.00 kW
(0.01 kW)
720.0 kW
(0.1 kW)
The unit [W] is replaced by [VA] for apparent power values.
Balanced
3‑phase/3‑wire
reactive power
(
Q
(3P3W)
)
Accuracy
±1 dgt relative to calculation from measured values
Range
configuration
For the above active power range configuration
,
the unit [W] is replaced
by [var].
HIOKI CM3286C961-00