84
Accuracy Table
AC current peak
value (
I
|PEAK|
)
Zero to Peak
No polarity
(absolute value
of the maximum
wave height
during the
display update
interval)
Range
(Accuracy guarantee
range is specified in
terms of current RMS
values.)
Resolution
Accuracy
Display range
45 Hz ≤ f
≤ 66 Hz
66 Hz < f
≤ 500 Hz
500 Hz < f
≤ 1 kHz
6.000 A
(0.060 A to 6.000 A)
0.01 A
±3.0% rdg ±5 dgt
±5.0% rdg
±5 dgt
0.00 A to 18.00 A
60.00 A
(0.60 A to 60.00 A)
0.1 A
±2.5% rdg ±5 dgt
±4.0% rdg
±5 dgt
0.0 A to 180.0 A
600.0 A
(6.0 A to 600.0 A)
1 A
–
0 A to 1000 A
AC current
frequency
(
FREQ
I
)
Range
(Accuracy guarantee
range)
Resolution
Accuracy
Current frequency values are shown as
“–
–
–
–” when the current RMS value is
less than 150 count. Current frequency
values of less than 45.0 Hz are shown
as “–
–
–
–.”
Display range
999.9 Hz
(45.0 Hz to 999.9 Hz)
0.1 Hz
±0.3% rdg ±3 dgt
45.0 Hz to 999.9 Hz