5.3 Timing Chart
81
5
Chap
ter
5
Ex
ter
nal
Cont
rol
(EXT
I
/O)
External Trigger Timing Chart
Internal Trigger Timing Chart
5.3 Timing Chart
Contact State
*1 ERR Output
(Err Output ASYNC Setting)
*3 TRIG Input
Measurement start signal
INDEX Output
Reference Signal
*4 EOM Output
End-of-Measurement Signal
Comparator Result
*2 ERR Output
(Err Output SYNC Setting)
Contact
Open
End of Conversion
Measuring
*1: For details, see “"ERR Output"(
p.79).”
*2: When ERR output is set to the SynChronous mode, measurement fault detection results can be obtained
when measurement is finished, as with comparator results.
*3: After connecting to the test object, wait for longer than the response time (approximately 700 ms) before
inputting the TRIG signal (It is necessary to wait out the response time for the measurement values to sta-
bilize after connection. Response times depend on the test object).
*4: When the EOM signal is set to pulse output, the signal will only turn on for the specified time after conver-
sion is complete.
t1
t1
t2
t3
t4
t5
t6
ON
OFF
ON
OFF
OFF
ON
OFF
ON
HOLD
Pulse
HOLD
Pulse
Measuring
Measuring
t3
t4
t5
* INDEX Output
Reference Signal
* EOM Output
End-of-Measurement
Signal
Comparator Result
* When the EOM signal is set to PULSE, the signal will remain on only for the specified period upon com-
pletion of conversion.
t3
OFF
ON
ON
OFF
t6
t6
HOLD
Pulse
HOLD
Pulse
HIOKI BT3564A961-02
Summary of Contents for BT3564
Page 2: ...HIOKI BT3564A961 02...
Page 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Page 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Page 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Page 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Page 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Page 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Page 206: ...Index 4 HIOKI BT3564A961 02...
Page 207: ...HIOKI BT3564A961 02...
Page 208: ...HIOKI BT3564A961 02...
Page 209: ...HIOKI BT3564A961 02...
Page 210: ...HIOKI BT3564A961 02...