8.6 Message Reference
129
8
Chapt
er 8
RS-232
C
/GP-I
B
In
te
rf
aces
Set and Query the Comparator Lower Threshold Setting
(Resistance Measurement)
Syntax
Command
:CALCulate:LIMit:RESistance:LOWer
<Lower threshold>
Query
:CALCulate:LIMit:RESistance:LOWer?
Response
<Lower threshold>
<Lower threshold>
= 0 to 99999
(NR1)
Example
Command
:CALC:LIM:RES:LOW 28406
Sets the lower threshold to 284.06 m
(with the 300 m
range
selected)
(If the 3
range is selected, the threshold is set to 2.8406
)
Query
:CALC:LIM:RES:LOW?
Response
28406
Note
The value is sent as a whole integer (count). To set 120.53 m
with the 300 m
range, send the following:
:CALC:LIM:RES:LOW 12053
(Voltage Measurement)
Syntax
Command
:CALCulate:LIMit:VOLTage:LOWer
<Lower threshold>
Query
:CALCulate:LIMit:VOLTage:LOWer?
Response
<Lower threshold>
<Lower threshold>
= 0 to 999999
(NR1)
Example
Command
:CALC:LIM:VOLT:LOW 360000
Sets the lower threshold to 3.60000 V
(with the 10 V range selected)
Query
:CALC:LIM:VOLT:LOW?
Response
360000
Note
The value is sent as a whole integer (count).
To set 45.9997 V with the 100 V range, send the following:
:CALC:LIM:VOLT:LOW 459997
HIOKI BT3564A961-02
Summary of Contents for BT3564
Page 2: ...HIOKI BT3564A961 02...
Page 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Page 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Page 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Page 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Page 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Page 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Page 206: ...Index 4 HIOKI BT3564A961 02...
Page 207: ...HIOKI BT3564A961 02...
Page 208: ...HIOKI BT3564A961 02...
Page 209: ...HIOKI BT3564A961 02...
Page 210: ...HIOKI BT3564A961 02...